Natural Convection Oven (General)
New type of natural convection oven that does not scatter the sample
Model Name : ON3-05/10/15, 05W/10W/15W
Superior performance
Wide temperature range that can be controlled up to 250°C.
Gravity flow convection for uniform sample heating.
- Ideal for all sterilization, drying, and high-temperature storage tasks that do not require high drying performance or airflow.
Excellent temperature stability maintains fluctuation within 0.2°C at 100°C
Superior insulation and enclosed design minimize energy loss and reduce operating costs.
Stainless steel interior and shelves are excellent in terms of corrosion resistance and clean maintenance.
Effective heat transfer through airflow
Structural functional features
Oven in which the air heated by the heater located on the inner bottom side is naturally air flows on the inside.
Easy to open and close the door. Soft handle minimizes shock to the sample.
Includes vent hole base with cover for gas exhaust.
Excellent visibility and viewing angle due to application of 4.7-inch large VFD.
Use Convenience Feature
3-point temperature calibration, high temperature accuracy over a wide temperature range.
Optimized control with temperature auto-tuning.
Save and use 3 frequently used temperatures.
Wait On/Off timer. (up to 999 hours 59 minutes)
Microprocessor PID method for precise temperature control.
Monitor via mobile app anytime, anywhere with LC Connected mobile monitoring system
(when purchased LC GreenBox)
Automatic restart of operation when power is restored after sudden power failure.
Over temperature limit function. (Over Temperature Limiter)
Safe product surface, even when operated at upper temperature.
Triple-toughened viewing window for safe and convenient observation even at high temperatures. (ON3-W)
Malfunction prevented by controller lock function.
Self-diagnostics function to check abnormal condition of temperature sensor.
Excellent visibility with large 4.7-inch VFD
Hands-free door opening
2-level loading with load accessories.
Model
Solid Door (without window)
ON3-03
ON3-05
ON3-10
ON3-15
Window door
ON3-03W
ON3-05W
ON3-10W
ON3-15W
Chamber volume (L / cu ft)
30 / 66.1
50 / 110.2
100 / 220.5
163 / 359.4
Temperature
Range (℃ / ℉)
Amb. +15 to 250
/ Amb. +27 to 482
Amb. +15 to 250
/ Amb. +27 to 482
Amb. +15 to 250
/ Amb. +27 to 482
Amb. +15 to 250
/ Amb. +27 to 482
Fluctuation at 100℃ (±℃ / ℉)
0.3 / 0.54
0.3 / 0.54
0.3 / 0.54
0.5 / 0.9
Variation at 100℃ (±℃ / ℉)
2.7 / 4.86
3.7 / 6.66
7.5 / 13.5
4.1 / 7.38
Heating time to 100℃ (min.)
78
67
74
61
Recovery time at 100℃ (min.)
18
34
39
35
Dimension
(W×D×H)
Interior (W x D x H, mm / inch)
310 x 380 x 252
/ 12.2 x 15.0 x 9.9
367 x 437 x 309
/ 14.4 x 17.2 x 12.2
463 x 533 x 405
/ 18.2 x 21.0 x 15.9
550 x 546 x 542
/ 21.7 x 21.5 x 21.3
Exterior (W x D x H, mm / inch)
488 x 643 x 503
/ 19.2 x 25.3 x 19.8
545 x 699 x 560
/ 21.5 x 27.5 x 22.0
641 x 796 x 655
/ 25.2 x 31.3 x 25.8
728 x 815 x 793
/ 28.7 x 32.9 x 31.2
Net weight (Kg / lbs)
43 / 94.8
53 / 116.84
63 / 138.9
73 / 16.9
Shelves
Quantity of shelves (standard/max.)
2 / 2
2 / 4
2 / 5
2 /7
Max. Load per shelf (Kg / lbs)
30 / 66.1
30 / 66.1
30 / 66.1
30 / 66.1
Electrical requirements (230V, 50 / 60Hz, A)
3.3
3.6
5.7
7.3
Cat. No.
Solid Door (without window)
AAH11165K
AAH11175K
AAH11185K
AAH11195K
Window door
AAH11265K
AAH11275K
AAH11285K
AAH11295K
Electrical requirements (120V, 60Hz, A)
6
6.7
10.8
14.1
Cat. No.
Solid Door (without window)
AAH11166U
AAH11176U
AAH11186U
AAH11196U
Window door
AAH11266U
AAH11276U
AAH11286U
AAH11296U
※ The lowest temperature that can be controlled depends on changes in room temperature and sample temperature. Therefore, please contact the distributor of our products before purchasing for technical consultation.
※ Technical data (according to DIN 12880, after 2013)