I. Preface
Olive oil is directly cold-pressed from fresh olive fruits without heating or chemical treatment, thus retaining its natural nutrients. Olive oil is regarded as the most suitable oil for human consumption among all the oils discovered so far. It is hailed as liquid gold in the West and has excellent natural health benefits, beauty effects and ideal cooking uses.
This method employs fully automatic potentiometric titration to determine the acid value of olive oil. It features good repeatability and significant jumps, enabling accurate measurement of the acid value of the oil and providing a precise basis for oil product testing.
Ii. Instruments and Reagents
2.1 Instruments
JH-T7 fully automatic potentiometric titrator, PH composite electrode, analytical balance, etc
2.2. Reagents
Ether - isopropanol (1:1) solution, sodium hydroxide (0.1mol/L) standard solution.
Iii. Experimental Methods
3.1 Sample testing
Accurately transfer 20g of the oil sample into the titration cup, add 50ml of the ether - isopropanol (1:1) mixed solution, mix evenly, place the sample on the potentiometric titration stand, start stirring, insert the electrode and titration head, and after the potential stabilizes, titrate to the end point with sodium hydroxide (0.1mol/L) standard solution, and record the end titration volume. The parameters of the titrator are set as shown in Table 1
Jiahang Table 1 Titration Parameter Settings
Titration mode: | Dynamic titration | stirring speed |
0.04mL |
Electrode equilibrium time |
4s |
pre-stirring time : |
10s |
Electrode equilibrium potential: |
1mv |
titration speed: | Standard |
Minimum addition volume: |
10mL |
Pre-titration addition volume |
56.1 |
Final volume: |
100 |
Pre-titration stirring time: |
400 |
Equilibrium potential before titration: |
6 |
liquid filling rate: |
6mv |
Pre-controlled pH value: | |||
Potential spike: |
3.3 Test Graph Example
Iv. Results and Discussion
4.1 Experimental Results
The experimental results are shown in Table 2
Inspection items |
Sample name |
Sample quality /g |
Titrant concentration (hydrochloric acid) /(mol/L) |
Titrator volume /mL |
Acid value (mgKOH/g |
Average value |
Acid value |
blank |
/ |
0.10414 |
0.120 |
/ |
/ |
1 |
20.38 |
4.093 |
0.5633 |
0.5739 |
||
2 |
20.08 |
4.124 |
0.5761 |
|||
3 |
21.98 |
4.563 |
0.5823 |
4.2、Conclusion
The acid value of olive oil tested in this test is 0.5739mgKOH/g. The data has good repeatability and the operation is safe and simple. It is a good choice for detecting acid value.